The DDM 1000 is designed for precise temperature measurements as demanded in calibration-laboratory use and quality assurance. The accuracy is adapted to the uncertainties of fix-point-cells for precise sensor calibrations. Pt 10-, Pt-25- and Pt 100-Sensors can be used at the same time. The resolution of up to 0,02 mK often is needed in physical […]
The DDM 1000 is designed for precise temperature measurements as demanded in calibration-laboratory use and quality assurance. The accuracy is adapted to the uncertainties of fix-point-cells for precise sensor calibrations. Pt 10-, Pt-25- and Pt 100-Sensors can be used at the same time. The resolution of up to 0,02 mK often is needed in physical and chemical laboratories.
Product features
Technical data | The following specifications are only valid for a nominal operating voltage and a environmental temperature of 23°C. |
Measuring range | -200°C up to 962°C (according to ITS-90) -200°C up to 850°C (according to DIN EN60751) |
Resolution | 0,02 mK |
Measuring uncertainty | 0,1 mK (optionally 5 mK in the range of -50°C up to 250°C) |
Measuring channels | 2 (optionally up to 81 with external multiplexers) |
Sensors | Pt-100/Pt-25/Pt-10 in a four-conductor-technology |
Connectors | Lemo 1S, quadripolar, in front panel |
Measuring current | about 0,5 mA switched DC (about 1 mA for Pt-25) |
Measuring time | about 1 sec. per channel, about 0,1 sec. per channel with reduced resolution |
Long-term stability | 5 mK / year |
Temperature coeff. | 1 mK/K |
Display | LCD, LED backlight, 2 lines ¨¤ 16 characters, 9 mm height |
Units | °C, °F, K, Ohm |
Interface | RS-232, isolated |
Operation | menu orientated, alternatively controllable by your PC |
Memory | 128 KB (optionally 512 KB) |
Suppy | 230 V, 50 Hz / 115 V, 60 Hz about 15 VA |
Size | 260 x 80 x 240 (width x height x depth in mm) |
Weight | about 2,5 kg |
Copyright - 2024 - Hoskin Scientific