            Temperature range RT…450 °C: Dry block Infrared Block Ø 60 mm / depth 150 mm SIKA temperature calibrators can be used everywhere, including on site in workshops, test and measurement rooms as well as laboratories. They cover a broad spectrum of industries with diverse applications. Assembly and commissioning Manufacturing and production Maintenance and service Quality […]
SIKA temperature calibrators can be used everywhere, including on site in workshops, test and measurement rooms as well as laboratories. They cover a broad spectrum of industries with diverse applications.
	Dry-block function
	The optimum thermal coupling from the block to test item is achieved 
with the correct adapter sleeve. Ideally, the internal diameter of the 
sleeve is 0.5 mm larger than the outer diameter of the test item. With 
the aid of the adapter sleeve, straight temperature sensors of virtually
 any length and diameter can be calibrated. The dry block covers the 
entire temperature range without the need to change the calibration 
medium. Viscosity, flash points or outgassing are of no concern.
	Black body function
	A patented infrared calibration sleeve is used to calibrate IR 
pyrometers or thermal imaging cameras. The special surface structure and
 the asymmetrical shapes create a „cavity radiator“ with an emission 
factor of 0.9994, prevent the reflection of interference radiation and 
emit the required temperature in an ideal form. The pyrometer is simply 
held at the specified distance above the measurement opening, thereby 
forming the desired measurement area on the bottom for the calibration 
to be performed. A support base can be fitted directly on the unit.
| Dry block | |
| Temperature range | RT...450 °C | 
| Tolerance | ±0,3 °C | 
| Stability | ±0,05 °C | 
| Infrared | |
| Temperature range | RT...450 °C | 
| Tolerance | ±0,5 °C | 
| Stability | ±0,05 °C | 
| Resolution | 0,01 °C in the range of RT...99,99 °C, else 0,1 °C | 
| Block | Ø 60 mm / depth 150 mm | 