The GLS Sampler is designed for general purpose or priority pollutant sampling in municipal and industrial applications where a full-size sampler is too large.
Size (Diam. x Height): | 16.5 x 26.5 inches (42 x 67 cm) |
---|---|
Weight: | 24.5 lbs (11.1 kg) (dry, less battery) |
Sample bottles: | 1.0 or 2.5 gallon polyethylene or glass container. |
Power Requirements: | 12 volts DC. (Supplied by battery or AC power converter.) |
Intake purge: | Purge before and after each sample. |
---|---|
Tubing life indicator: | Provides a message to change pump tubing. |
Intake suction tubing: | Length 3 to 99 ft. (1 to 30 m), Vinyl or Teflon® lined, ID 0.375 in. or 0.250 in |
Pump tubing life: | 2,000 samples |
Typical repeatability: | ±10 ml |
Maximum lift: | 26 ft. (7.9 m) |
Typical line velocity @ Head height: | 2.9 ft./s @ 3 ft; 2.5 ft./s @ 10 ft; 1.9 ft./s @ 15 ft. (0.76 m/s @ 0.9 m; 0.76 m/s @ 3.1 m; 0.58 m/s @ 4.6 m) |
Sample presence detector: | Non-wetted, non-conductive sensor detects when liquid sample reaches the pump to automatically compensate for changes in head heights. |
Operational Temperature: | 32° to 120°F (0° to 49°C) |
Enclosure: | NEMA 4X, 6 IP67 |
---|---|
Program storage: | 1 sampling program (in non-volatile ROM) |
Flowmeter signal input: | 5 to 15 volt DC pulse or 25 millisecond contact closure. |
Number of composite samples: | Up to 500 maximum |
Clock accuracy: | 1 minute per month, typical, for Real Time Clock |
Sampling modes: | Uniform time, flow. (Flow mode is controlled by external flow meter pulses.) |
Sample frequency: | 1 minute to 9,999 minutes, in 1 minute increments. 1 to 9,999 flow pulses. |
Sample volumes: | Programmable 10 to 9,990 ml in 1 ml increments |
Controller diagnostics: | Tests for RAM, ROM, pump, and display |
Copyright - 2024 - Hoskin Scientific