Canada’s Instrumentation Leader Since 1946.

Petra MAX

Petra MAX complies with ASTM D4294, ISO 8754 and IP 336 methods for sulfur analysis of hydrocarbons like crude oil, diesel, gasoline, jet fuel, and lubricants.

  • SKU : G-F45-403102-01
  • Manufacturer: XOS
Contact us for details

Petra MAX delivers D4294 sulfur analysis in addition to 12 elements from P to Zn, for rapid monitoring of critical elements like Ca, Fe, K, Ni, and V at sub-ppm levels. Petra MAX offers advanced precision with HDXRF, advanced reliability, and advanced software and data management. Petroleum laboratories depend on reliable, robust analytical solutions for their fast-paced environment. Petra MAX was designed to meet these needs with an innovative sample introduction system that directs accidental spills to a drip tray and away from valuable components.

Petra MAX Autosampler offers a more efficient workflow with sample tracking and continuous sample loading. Users can eliminate data errors and add urgent samples to the queue as needed.

The Autosampler Upgrade Kit can be added to any Petra Series analyzer. Users have the option to use X-ID Sample Cups (QR-coded) or standard XRF cups.

Check out Petra MAX autosampler – delivering a more efficient workflow with sample tracking and continuous sample loading.

  • Advanced Sulfur Analysis with HDXRF
  • Advanced Precision
  • Advanced Applications
  • Advanced Software
  • Advanced Data Management

 

Details

Advanced Sulfur Analysis with EDXRF

Enhanced detection performance over traditional XRF technology

Advanced Precision

Precise sulfur measurements with a LOD as low as 5.7 ppm and rapid monitoring of critical elements at sub-ppm levels

Advanced Applications

Hydrocarbons, water and catalysts

Advanced Software

Preset your measurement configurations

Advanced Data Management

Storage for thousands of measurements

 

Specifications
Auto Sampler: No
Calibration curves: 30 calibration curves. Linear (automatic calibration available)
Catalysts: Yes
Communication Capabilities: Printer, USB & Ethernet, LIMS Integrated
Consumable/Utility Gas: None
Dimensions: Depth: 16.5 in / 41.9 cm
Dimensions: Height: 6 in / 15.3 cm
Dimensions: Width: 14.5 in / 36.8 cm
Element: Sulfur / 5.7 ppm
Phosphorus / 17 ppm
Chlorine / 3 ppm
Potassium / 0.7 ppm
Calcium / 0.4 ppm
Vanadium / 0.1 ppm
Chromium / 0.09 ppm
Manganese / 0.07 ppm
Iron / 0.07 ppm
Cobalt / 0.07 ppm
Nickel / 0.04 ppm
Copper / 0.1 ppm
Zinc / 0.1 ppm
Measurement Time: User selectable 30-900 s
Method Name: ASTM D4294, ISO 8754 & IP 336
Options: Vibration Plate
Range: Sulfur 5.7 ppm – 10 wt%
Sample Cup Type: Standard XRF Sample Cup
Sample Matrix: Hydrocarbon
Specific Technology: EDXRF
Type: Benchtop