The ST2 is similar to ST1, except that it is fitted with a higher specified thermistor sensor whose accuracy over a temperature range of 0 to 70°C (+/- 0.1°C) Thermistor – contained in a stainless steel sheath of 4.8mm nominal outside diameter, 125mm long, which is, in turn, connected to a two core cable, 5m […]
Type |
Probe Type |
Probe accuracy (°C) |
Probe range (°C) |
Cable type |
Number of cores |
Cable range (°C) |
Logger accuracy |
|||
DL2e
(a) (°C) |
DL6(b) (°C) |
GP1(c) (°C) |
GP2(d) (°C) |
|||||||
2
k |
±0.2 |
‐50
to 150 |
PVC |
2 |
‐10 to +105 |
±0.14 |
±0.4 |
±0.4 |
±0.09 |
|
2
k |
±0.1(f) |
‐50
to 150 |
PVC |
2 |
‐10
to +105 |
±0.14 |
±0.4 |
±0.4 |
±0.09 |
|
2
k |
±0.2 |
‐20 to +80 |
Poly‐
ethelene |
2 |
‐20 to +80 |
±0.14 |
±0.4 |
±0.4 |
±0.09 |
|
10
k |
±0.2 |
‐10 to +40 |
PVC |
2(e) |
‐10 to +40 |
Not
available |
Not
available |
±0.14 |
±0.08 |
|
k |
±1.5 |
‐50
to 250 |
PTFE |
2 |
‐50 to +250 |
±1.0 |
Not
available |
Not
available |
±1.0 |
|
2
k |
±0.2 |
‐50
to 150 |
PVC |
2 |
‐10
to +105 |
±0.14 |
±0.4 |
±0.4 |
±0.09 |
|
2
k |
±0.1(f) |
‐50
to 150 |
Vy/TEFL(g) |
4 |
Not available |
±0.14 |
±0.4 |
±0.4 |
±0.09 |
|
2
k |
±0.1(f) |
‐50
to 150 |
Nylon
‐
PVC |
2 |
‐10 to +105 |
±0.14 |
±0.4 |
±0.4 |
±0.09 |
|
2
k |
±0.1(f) |
‐50
to 150 |
PVC |
2 |
‐10
to +105 |
±0.14 |
±0.4 |
±0.4 |
±0.09 |