Temperature range -35…165 °C: Micro bath Dry block Infrared -25…150 °C: Surface Block Ø 60 mm / depth 170 mm SIKA temperature calibrators can be used everywhere, including on site in workshops, test and measurement rooms as well as laboratories. They cover a broad spectrum of industries with diverse applications. Assembly and commissioning Manufacturing and […]
SIKA temperature calibrators can be used everywhere, including on site in workshops, test and measurement rooms as well as laboratories. They cover a broad spectrum of industries with diverse applications.
Micro bath function
The use of calibration liquids offers certain advantages if temperature
sensors with an unusual shape and size are to be tested. The test item
is immersed directly into the liquid without an insulating air gap,
resulting in direct temperature contact between the calibrator and the
test item. The liquid, such as silicone oil, is chosen depending on the
calibration temperature required. The continuous adjustment of the
magnetic stirrer together with the removable sensor basket agitates the
calibration liquid to create a large measuring zone. Furthermore, the
sensor basket guarantees unhindered stirring and helps protect the tank
floor.
Dry-block function
The optimum thermal coupling from the block to test item is achieved
with the correct adapter sleeve. Ideally, the internal diameter of the
sleeve is 0.5 mm larger than the outer diameter of the test item. With
the aid of the adapter sleeve, straight temperature sensors of virtually
any length and diameter can be calibrated. The dry block covers the
entire temperature range without the need to change the calibration
medium. Viscosity, flash points or outgassing are of no concern.
Black body function
A patented infrared calibration sleeve is used to calibrate IR
pyrometers or thermal imaging cameras. The special surface structure and
the asymmetrical shapes create a „cavity radiator“ with an emission
factor of 0.9994, prevent the reflection of interference radiation and
emit the required temperature in an ideal form. The pyrometer is simply
held at the specified distance above the measurement opening, thereby
forming the desired measurement area on the bottom for the calibration
to be performed. A support base can be fitted directly on the unit.
Surface temperature function
Surface temperature sensors are calibrated using special sleeves that
are fitted vertically with the required contact force. Switching
calibration control to the external reference sensor creates the best
possible temperature reference point on the surface of the sleeve. The
reference sensor is located directly beneath the abutting face of the
sleeve. The sleeve is designed in such a way that the best temperature
homogeneity is achieved in the centre of the abutting face. The special
design of the abutting face enables good thermal contact. There is no
need to use a thermally conductive paste or other thermal conduction
aids.
Micro bath | |
Temperature range | -35...165 °C |
Tolerance | ±0.1 °C |
Stability | ±0.05 °C |
Dry block | |
Temperature range | -35...165 °C |
Tolerance | ±0.3 °C |
Stability | ±0.05 °C |
Infrared | |
Temperature range | -35...165 °C |
Tolerance | ±0.5 °C |
Stability | ±0.05 °C |
Surface | |
Temperature range | -25...150°C |
Tolerance | ±1 °C |
Stability | ±0.2 °C |
Resolution | 0.1 / 0.01 / 0.001 °C |
Block | Ø 60 mm / depth 170 mm |
Copyright - 2024 - Hoskin Scientific