Ultrasonic Setting Measurement

The early setting is usual measured with the penetrometer or the Vicat apparatus. This kind of measurement has some disadvantages. In the early sixties some researchers suggested to use the ultra­sonic transition time as method for measuring the setting of mortars or fresh concrete. In cooperation with a German dry mix company, Schleibinger developed a […]

  • SKU : G-I04-U0001 VicaSonic
Ask us about this item
The early setting is usual measured with the penetrometer or the Vicat apparatus. This kind of measurement has some disadvantages. In the early sixties some researchers suggested to use the ultra­sonic transition time as method for measuring the setting of mortars or fresh concrete. In cooperation with a German dry mix company, Schleibinger developed a special ultrasonic unit with an integrated datalogger and a special designed setup for the mortar specimen.

Measurement Principle

Fresh mortar is setting between two ultrasonic transducers, which are supplied with 0.25s or less pulses per second at a frequency of 54 kHz. The mortar changes, according to the setting, the ultrasonic transition time. From the beginning of first contact of the growing binder-stone, the ultrasonic sound speed increases. If there is a complete crystalline structure the change of the sound speed will be slower. Until the a further aggregation of the cement stone it increases again.

Measurement Setup

Each measurement cell is supplied with a pair of ultrasonic transducers. The Vikasonic continuously measures the ultrasonic transition time and the signal damping. The data sets are recorded digitally to a common USB stick. Also the sound speed and the elastic modulus is calculated and recorded. Additionally the temperature inside the specimen is measured with a type K thermocouple.