Canada’s Instrumentation Leader Since 1946.

UOP10 – Laboratory steam generator

As many laboratories do not incorporate a centralised or reliable steam source, Armfield offers an advanced boiler providing a constant supply of steam. The compact unit requires only connection to water and electrical supplies for operation.

Contact us for details
Description

As many laboratories do not incorporate a centralised or reliable steam source, Armfield offers an advanced boiler providing a constant supply of steam. The compact unit requires only connection to water and electrical supplies for operation.

The Armfield UOP10 Laboratory Steam Generator comprises a mains fed header tank with level control, which supplies the pressure vessel via a quiet running, heavy duty, high pressure feed pump.

The pressure vessel is built to all relevant British and European standards. Being heavily insulated and featuring two totally immersed Incoloy heating elements, it provides a highly efficient and economical means of generating a continuous supply of steam.

The two heating elements are individually switched and the unit will operate satisfactorily on a single element.

The integral control panel houses an on/off mains switch with switchable test functions, LEDs display the operational status of the equipment.

The UOP10 is fully safeguarded via a Magnetic Circuit Breaker (MCB) and a Residual Current Device (RCD) to give a long trouble free life.

The components of the UOP10 are mounted within a steel cabinet finished with a high gloss stove enamel paint for corrosion resistance and long life. Front panels are hinged for easy access and are fitted with key operated handles.

It is recommended that for optimum life of the unit, feedwater should be demineralised.

 

Features & Benefits
  • Simple controls allow for quick start up and shut down procedures
  • Variable pressure outputs
  • Rapid start up to full pressure and will deliver steam during heating process
  • Automatic top up of feed tank and boiler during normal use
  • Normal domestic water supply sufficient for normal operation
  • Low water and high pressure alarms cut power to heating element to prevent unsafe conditions
  • Dual locks to prevent unauthorised access to boiler internal controls
  • Lock out isolation switch on front of unit for safety during maintenance
  • Rapid and variable blowdown of steam boiler during shutdown of system
  • Pressure relief valves in case of over-pressurisation to prevent energetic discharge

 

Requirements
  • Electrical supply: Three phase (see ordering codes)
  • Water supply: 1.0l/min @ 2 bar maximum usage

 

Technical Specifications

Steam output at 100˚C – 55Kg/hr (continuous)

Heat output – 36kW

Number of heating elements – 2

Max. operating pressure – 10 bar

 

Dimensions

Length: 1.12m

Width: 0.68m

Height: 1.92m

Note: Total height of 2.5m is required for access