Device to monitor the radial profile of the specimen whilst under test using highly accurate laser precision. The lasers are installed on to a modified conventional triaxial cell and monitor a 32mm length on the sample from a distance of 80 mm. The laser heads are integrated into GDSLAB software. The software allows the laser […]
Device to monitor the radial profile of the specimen whilst under test using highly accurate laser precision. The lasers are installed on to a modified conventional triaxial cell and monitor a 32mm length on the sample from a distance of 80 mm.
The laser heads are integrated into GDSLAB software. The software allows the laser to measure a point measurement and also scan the sample profile, which may be used to give a measurement volume change. GDSLAB can record a profile scan of every 5 seconds. The laser profilers have an accuracy of about ± 10μm in optimal conditions of measurement.
Need advice choosing the right transducer, view our options table here.