A transducer sensor that can measure either air pressure or velocity with the flip of a switch. Available in the universal configuration, with pressure and velocity range options: 0-10″ WC / 0-7,000 ft/min, with seven field-selectable sub-ranges for pressure and eight for velocity. Available with display. This sensor has an IP65/NEMA 4 environmental rating.

Requirements for use with specific data loggers:

  • HOBO MX1100 series: a 0-5 Volt DC input cable (SD-VOLT-05) and external power provided by an AC adapter (AC-SENS-1)
  • HOBO U12, UX120-006M, or ZW series: a 0-5 Volt DC input cable (CABLE-ADAP5) and external power provided by an AC adapter (AC-SENS-1)
  • HOBO RX3000: an analog input module (RXMOD-A1)
  • HOBO U30: an analog input module (VIA)
  • HOBO H22: an analog input module (S-FS-CVIA)

Highlighted Features

  • Switch-selectable ranges and scales reduce setup time
  • High reliability sensor technology for long-term maintenance-free operation
  • Monitors positive and negative pressure for application versatility

This sensor operates in an indoor environment.

The T-VER-PX3UL sensor supports the following measurement: Differential Pressure

  • Media Compatibility Dry air or inert gas
  • Input Power 12-30VDC, or 24VAC nominal
  • Output Field-selectable: 2-wire, loop-powered 4-20mA (DC only, clipped and capped) or 3-wire 0-5V/0-10V
  • Pressure Ranges: Unidirectional: 0.1/0.25/0.5/1.0” W.C. F.S. (25
    pa/50 pa/100 pa/250 pa, F.S.), switch selectable; Bidirectional:
    ±0.1/±0.25/±0.5/±1.0” W.C. F.S. (±25 pa/±50 pa/±100 pa/±250 pa, F.S.),
    switch selectable
  • Response Time Standard: T95 in 20 sec, Fast: T95 in 2 sec, switch selectable
  • Mode Unidirectional or bidirectional, switch selectable
  • Display (option) Signed 3-1/2 digit LCD, indicates pressure, overrange indicator
  • Proof Pressure 3 psid (20.6kPa)
  • Burst Pressure 5 psid (34.5kPa)
  • Accuracy ±1% F.S. of selected range (combined linearity and hysteresis)
  • Temperature Effect 1″ 0.05%/°C; 10″
  • Zero Drift (1-year) 1″ 2.0% max.
  • Zero Adjust Push button auto-zero and digital input (2-pos terminal block)
  • Operating Environment 0°- 60°C (32° to 140°F); 0 to 90% RH non-condensing
  • Fittings Brass barb; 0.24” (6.1mm) o.d.
  • Physical UL 94 V-O Fire Retardant ABS