Canada’s Instrumentation Leader Since 1946.

HOBO Water Level Data Logger (0 to 4 m)

$956.00

The HOBO Water Level data logger features high accuracy at a great price and ease-of-use, with no cumbersome vent tubes or desiccants to maintain. This data logger is ideal for recording water levels and temperatures in shallow wells, streams, lakes and freshwater wetlands.  For saltwater deployments, such as brackish wetlands and tidal areas, see HOBO U20 Water Level Titanium.

Contact us for details
Important Info

Overview

The HOBO Water Level data logger features high accuracy at a great price and ease-of-use, with no cumbersome vent tubes or desiccants to maintain. This data logger is ideal for recording water levels and temperatures in shallow wells, streams, lakes and freshwater wetlands.  For saltwater deployments, such as brackish wetlands and tidal areas, see HOBO U20 Water Level Titanium.

Highlighted Features

  • Lightning protection – no long signal wires, and electronics are shielded in stainless steel housing for use in freshwater
  • HOBOware Pro software provides easy conversion to accurate water level reading, fully compensated for barometric pressure temperature, and water density.
  • Multiple-rate sampling allows faster sampling at critical times such as when pumping starts or stops.
  • Available in 4 depth ranges
  • Ideal for use in wells, streams, lakes, wetlands and tidal areas
  • No-vent-tube design for easy reliable deployment
  • Available in stainless and titanium versions
  • Durable ceramic pressure sensor
  • 3-point NIST-traceable calibration certificate included

In what environment does this data logger operate?

This data logger operates in outdoor and underwater environments.

What measurements does this data logger support?

The U20-001-01 data logger supports the following measurements: Absolute Pressure, Barometric Pressure, Temperature and Water Level

Specifications
Pressure (Absolute) and Water Level Measurements MX2001-01-S and MX2001-01-Ti-S – 9 Meter (30′) range​
Operation Range: 0 to 207 kPa (0 to 30 psia); approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, or 0 to 12 m (0 to 40 ft) of water at 3,000 m (10,000 ft) of altitude
Factory Calibrated Range : 69 to 207 kPa (10 to 30 psia), 0° to 40°C (32° to 104°F)
Burst Pressure : 310 kPa (45 psia) or 18 m (60 ft) depth
Water Level Accuracy* : Typical error: ±0.05% FS, 0.5 cm (0.015 ft) water
Maximum error: ±0.1% FS, 1.0 cm (0.03 ft) water
Raw Pressure Accuracy** : ±0.3% FS, 0.62 kPa (0.09 psi) maximum error
Resolution: < 0.02 kPa (0.003 psi), 0.21 cm (0.007 ft) water
Pressure Response Time (90%)***: 1 second at a stable temperature
Pressure (Absolute) and Water Level Measurements MX2001-02-S – 30 Meter (100′) range​
Operation Range : 0 to 400 kPa (0 to 58 psia); approximately 0 to 30.6 m (0 to 100 ft) of water depth at sea level, or 0 to 33.6 m (0 to 111 ft) of water at 3,000 m (10,000 ft) of altitude
Factory Calibrated Range : 69 to 400 kPa (10 to 58 psia), 0° to 40°C (32° to 104°F)
Burst Pressure: 500 kPa (72.5 psia) or 40.8 m (134 ft) depth
Water Level Accuracy*: Typical error: ±0.05% FS, 1.5 cm (0.05 ft) water
Maximum error: ±0.1% FS, 3.0 cm (0.1 ft) water
Raw Pressure Accuracy**: ±0.3% FS, 1.20 kPa (0.17 psi) maximum error
Resolution: 0.04 kPa (0.006 psi), 0.41 cm (0.013 ft) water
Pressure Response Time (90%)***: 1 second at a stable temperature
Pressure (Absolute) and Water Level Measurements MX2001-03-S – 76 Meter (250′) range
Operation Range: 0 to 850 kPa (0 to 123.3 psia); approximately 0 to 76.5 m (0 to 251 ft) of water depth at sea level, or 0 to 79.5 m (0 to 262 ft) of water at 3,000 m (10,000 ft) of altitude
Factory Calibrated Range : 69 to 850 kPa (10 to 123.3 psia), 0° to 40°C (32° to 104°F)
Burst Pressure: 1,200 kPa (174 psia) or 112 m (368 ft) depth
Water Level Accuracy* : Typical error: ±0.05% FS, 3.8 cm (0.125 ft) water
Maximum error: ±0.1% FS, 7.6 cm (0.25 ft) water
Raw Pressure Accuracy**: ±0.3% FS, 2.55 kPa (0.37 psi) maximum error
Resolution: < 0.085 kPa (0.012 psi), 0.87 cm (0.028 ft) water
Pressure Response Time (90%)***: 1 second at a stable temperature
Pressure (Absolute) and Water Level Measurements MX2001-04-S and MX2001-04-Ti-S – 4 Meter (13′) range​
Operation Range: 0 to 145 kPa (0 to 21 psia); approximately 0 to 4 m (0 to 13 ft) of water depth at sea level, or 0 to 7 m (0 to 23 ft) of water at 3,000 m (10,000 ft) of altitude
Factory Calibrated Range: 69 to 145 kPa (10 to 21 psia), 0° to 40°C (32° to 104°F)
Burst Pressure: 310 kPa (45 psia) or 18 m (60 ft) depth
Water Level Accuracy*: Typical error: ±0.075% FS, 0.3 cm (0.01 ft) water
Maximum error: ±0.15% FS, 0.6 cm (0.02 ft) water
Raw Pressure Accuracy**: ±0.3% FS, 0.43 kPa (0.063 psi) maximum error
Resolution: < 0.014 kPa (0.002 psi), 0.14 cm (0.005 ft) water
Pressure Response Time (90%)***: < 1 second at a stable temperature
Barometric Pressure (MX2001-TOP)
Operation and Calibrated Range: 66 to 107 kPa (9.57 to 15.52 psia); -20° to 50°C (-4° to 122°F)
Accuracy: ±0.2 kPa (±0.029 psi) over full temperature range at fixed pressure; maximum error ±0.5% FS
Water Level Accuracy*: Typical error: ±0.075% FS, 0.3 cm (0.01 ft) water
Maximum error: ±0.15% FS, 0.6 cm (0.02 ft) water
Resolution: 0.01 kPa (0.0015 psi)
Response Time: 1 second at stable temperature
Stability (Drift) : 0.01 kPa (0.0015 psi) per year
Temperature Measurements (All Sensor End Models MX2001-0x-S and MX2001-0x-Ti-S)
Operation Range: -20° to 50°C (-4° to 122°F)
Accuracy: ±0.44°C from 0° to 50°C (±0.79°F from 32° to 122°F)
Resolution: 0.1°C at 25°C (0.18°F at 77°F)
Response Time (90%) : 5 minutes in water (typical)
Stability (Drift) : 0.1°C (0.18°F) per year
Logger
Operation Range: -20° to 50°C (-4° to 122°F)
Radio Power : 1 mW (0 dBm)
Transmission Range: Approximately 30.5 m (100 ft) line-of-sight
Wireless Data Standard: Bluetooth Low Energy (Bluetooth Smart)
Logging Rate: 1 second to 18 hours
Logging Modes : Fixed interval, multiple intervals with up to 8 user-defined logging intervals and durations, or event-triggered burst
Memory Modes : Wrap when full or stop when full
Start Modes : Immediate, date & time, or next interval
Stop Modes : When memory full, stop with HOBOmobile, date & time, or after a set logging period
Time Accuracy : ±1 minute per month 0° to 50°C (32° to 122°F)
Battery: Two AA, 1.5 V alkaline batteries, factory-replaceable
Battery Life : 1 year, typical with logging interval of 1 minute. Faster logging and/or statistics sampling intervals, entering burst logging mode,
excessive readouts, checking of Full Status Details, and remaining connected with HOBOmobile will impact battery life.
Memory: 256 KB memory (30,000 sets of measurements)
Full Memory Download Time: Approximately 2 minutes; may take longer the further the device is from the top end of the logger
Dimensions: Top end (MX2001-TOP): 2.54 cm (1.0 inches) diameter, 28.9 cm (11.4 inches) length; mounting hole 7.6 mm (0.3 inches) diameter
Sensor end (MX2001-0x-S and MX2001-0x-Ti-S: 2.54 cm (1.0 inches) diameter, 9.91 cm (3.9 inches) length
Note: The length of the water level logger cable (CABLE-DR-xxx) can vary -0% to +3% +10 cm (3.9 inches) from the length ordered. The
logger adds 38.8 cm (15.3 inches) to the length of the cable ordered.
Weight: Top end (MX2001-TOP): Approximately 136 g (4.78 oz) in air
Stainless sensor end (MX2001-0x-S): Approximately 106 g (3.74 oz) in air; approximately 53.9 g (1.9 oz) in fresh water
Titanium sensor end (MX2001-0x-Ti-S): Approximately 80 g (2.83 oz) in air; approximately 37 g (1.3 oz) in fresh water
Wetted Materials: Top end (MX2001-TOP): Acetal housing, Polycarbonate end caps, Polycarbonate collar nut; Viton and Buna-N O-rings
Stainless sensor end (MX2001-0x-S): Acetal housing, PVC end cap, Polycarbonate collar nut; Viton and Buna-N O-rings; ceramic sensor in stainless steel end cap
Titanium sensor end (MX2001-0x-Ti-S): Acetal housing, PVC end cap, Polycarbonate collar nut; Viton and Buna-N O-rings; ceramic sensor in Titanium end cap
Environmental Rating: NEMA 6, IP67 (top end)
-The CE Marking identifies this product as complying with all relevant directives in the European Union (EU)

*Water Level Accuracy: With accurate reference water level measurement, known water density, and a stable temperature environment. System Water Level Accuracy equals the sum of the Barometric Water Level Accuracy plus the selected sensor end Water Level Accuracy.

**Raw Pressure Accuracy: Absolute pressure sensor accuracy includes all sensor drift, temperature, and hysteresis-induced errors.

***Changes in Temperature: Allow 20 minutes in water to achieve full temperature compensation of the pressure sensor. There can be up to 0.5% of additional error due to rapid temperature changes. Measurement accuracy also depends on temperature response time.

Email
Phone:
+1-800-665-5871
Our Offices